MyoPlus 2 Pro

2 channel Stimulator Amplitude: 0-90 mA into 500 Ohm load - actual mA will tend to be less than indicated due to Electrode impedance: at 1000 Ohms load (Electrodes in poor condition) the maximum will be limited to 75 mA, at 1500 Ohms load the maximum will be limited to 50 mA.

Type: Constant current, maximum output voltage 70 Volts +5 / -10 Volts

Waveform: Symmetrical, rectangular, bi-phasic with net zero DC current

Pulse width selection: 50-450 μS (2% accuracy)

Pulse rate selection: 2-100 Hz (2% accuracy)

Work / Rest periods: 2-99 seconds

Time 1 - 99 minutes

Ramp up time: 0.1 - 9.9 seconds

Preset and user programmable treatment Programs

Automatic output shut off with detection of open electrode above 0.5 mA

 